  • 测量原理

    Ultrasonic transit time difference measurement

  • Measured variables

    Mass flow rate, volumetric flow s. c. (standard condition), volumetric flow a. c. (actual condition), molecular weight, gas volume and mass, gas velocity, sound velocity

  • Measuring range

    0,03 m/s ... 120 m/s (0.1 ft/s ... 394 ft/s)

  • Max. Measurement error

    Volumetric flow, a. c.; 1 % ... 5 %
    Related to the measured value with ultrasonic technology (in the range ≥ 0.3 m/s (≥ 1 ft/s) to measuring range end value)
    0,5 % ... 1,5 % with spool piece and flow calibration
    Related to the measured value with ultrasonic technology (in the range ≥ 1.0 m/s (≥ 3.3 ft/s) to calibration range end value
     
    Mass flow rate a.c.; 2 % ... 5.5 %
    Related to the measured value with ultrasonic technology (in the range ≥ 0.3 m/s (≥ 1 ft/s) to measuring range end value)
    1,5 % ... 2 % with spool piece and flow calibration
    Related to the measured value with ultrasonic technology (in the range ≥ 1 m/s (≥ 3.3 ft/s) to calibration range end value)

  • Medium temperature range

    -196 °C ... +280 °C
    (-320 °F ... +536 °F)

  • Operating pressure range

    CL150 device flange (1) ATEX/IECEx: 20 bar(g)
    CSA: 16 bar(g)
    PN25 device flange (optional) (1) ATEX/IECEx: 20 bar(g)
    CSA: 16 bar(g)
    CL300 device flange (optional) (1) ATEX/IECEx: 20 bar(g)
    CSA: 16 bar(g)
    (1) Temperature dependent.

  • Nominal pipe size

    1-path measurement 4 ″ ... 86 ″
    2-path measurement 12 ″ ... 86 ″
    Other nominal sizes on request

  • Conformities

    ATEX: 2014/34/EU
    EMC: 2014/30/EU
    RoHS: 2011/65/EU
    PED: 2014/68/EU

  • Hazardous area approvals

    Interface Unit
    IECEx:
    Zone 1: Ex db eb ia IIC T4 Gb
    Zone 1: Ex db ia IIC T4 Gb
    Zone 2: EX ec ia IIC T4 Gc
     
    ATEX:
    Zone 1: II 2G Ex db eb ia IIC T4 Gb
    Zone 1: II 2G Ex db ia IIC T4 Gb
    Zone 2: II 3G Ex ec ia IIC T4 Gc
     
    NEC/CEC (US/CA):
    Class I, Zone 1, AEx d ia IIC T4 Gb
    Class I, Division 1, Groups B, C, D, T4
    Class I, Zone 2, AEx ec ia IIC T4 Gc
    Class I, Division 2, Group A, B, C and D, T4
     
    Sensors
    F1F-S/P/R:
    IECEx:
    Zone 1: Ex db [ia Ga] IIC T6 Ga/Gb
    ATEX:
    Zone 1: II 1/2 G Ex db [ia Ga] IIC T6 Ga/Gb
    NEC/CEC (US/CA):
    Class I, Division 1, Groups B, C and D, T4; 
    Class I, Zone 1, Ex/AEx d[ia] IIB + H2, T4;
     
    F1F-M:
    IECEx:
    Zone 1: Ex db [ia Ga] IIB T4 Ga/Gb
    ATEX:
    Zone 1:  II 1/2 G Ex db [ia Ga] IIB T4 Ga/Gb
    NEC/CEC (US/CA):
    Class I, Division 1, Groups C and D, T4; 
    Class I, Zone 1, Ex/AEx d[ia] IIB, T4;
     
    F1F-H:
    IECEx:
    Zone 1: Ex db IIC T6 Gb
    ATEX:
    Zone 1: II 2 G Ex db IIC T6 Gb
    NEC/CEC (US/CA):
    Class I, Division 1, Groups B, C and D, T4; 
    Class I, Zone 1, Ex/AEx d IIB + H2, T4;

  • Outputs

    Up to 6 outputs when using I/O modules (Option)
    16 bit:
    4 ... 20 mA
    7 V DC ... 30 V DC
    In accordance with NAMUR NE43
    Passive, galvanically isolated
    Reverse polarity protection

  • Digital Outputs

    Switching output:
    Up to 6 outputs when using I/O module (option)
    Electrically isolated
    Max. current: 70 mA
    Max. switching frequency: 50 Hz
    Max. input voltage: 30 V DC
    Max. saturation voltage at output: 0.5 V DC
    Switchable Namur/open collector
    Reverse polarity protection
     
    Digital output/Frequency output:
    Up to 12 digital outputs when using I/O modules (option)
    Electrically isolated
    Max. current: 50 mA
    Max. switching frequency: 10 kHz
    Max. input voltage: 30 V DC
    Max. saturation voltage at output: 1.8 V DC
    Switchable Namur/open collector
    Reverse polarity protection

  • Digital communication

    HART
    MODBUS
    FOUNDATION FIELDBUS